User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: HP 3456A:Digital Voltmeter DATE: 25-Aug-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 2 NUMBER OF LINES: 36 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON # This procedure is run with the MET/CAL Configuration #1 loaded. 1.001 ASK- R F W 1.002 ASK+ X 1.003 HEAD INITIAL CONDITIONS 1.004 DISP Allow a 1 hour warm-up for the UUT. 1.005 DISP Connect an IEEE cable from the System port 1 to 1.005 DISP the UUT HP-IB connector. 1.006 DISP Set the UUT front panel controls as follows: 1.006 DISP TERMINALS .................. FRONT 1.006 DISP GUARD ...................... TO LO 1.007 DISP Connect the 5700A and the UUT as follows: 1.007 DISP 5700A TO 3456A 1.007 DISP OUTPUT-HI ------------> VOLTS INPUT-HI 1.007 DISP OUTPUT-LO ------------> VOLTS INPUT-LO 1.008 CALL HP 3456A Sub Procedure 1 1.009 JMP 2.001 1.010 EVAL 2.001 CALL HP 3456A Sub Procedure 2 2.002 HEAD {} 2.003 END #! T.U.R.s less than 4.00: 8 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 2 #! T.U.R.s not computable at compile time: 0 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.